Knowledge Gateway
Unlock your learning potential with industry insights and standards that matter. Share your email to begin.
Unlock your learning potential with industry insights and standards that matter. Share your email to begin.
Standard Type : Main
Document No : ISO 27895:2009
Document Year : 2009
Pages : 11
Edition : 1
Section Volume : Vacuum technology (ISO/TC 112.)
ISO 27895:2009 specifies methods for the leak testing of vacuum valves used for control of gas flow or vacuum pressure in a vacuum system.It is applicable to vacuum valves that can be closed to leak rates less than 1 x 10-5 Pa m3/s for trace gas. The methods employ a sealing arrangement for the valve body, which is also specified in ISO 27895:2009. The methods are suitable for the verification of valve specifications.
A valve leak rate less than the nominal leak rate specified by the manufacturer during and after the operation enables the specification of such valve operating conditions as operating pressure range, permissible pressure difference between ports, bake-out temperature or operating temperature, and life cycle.