IEC TS 62558:2011

Ultrasonics - Real-time pulse-echo scanners - Phantom with cylindrical, artificial cysts in tissue-mimicking material and method for evaluation and periodic testing of 3D-distributions of void-detectability ratio (VDR)

Standard Details

IEC/TS 62558:2011(E) specifies essential characteristics of a phantom and method for the measurement of void-detectability ratio for medical ultrasound systems and related transducers. It is restricted to the aspect of long-term reproducibility of testing results. Medical diagnostic ultrasound systems and related transducers need periodic testing as the quality of medical decisions based on ultrasonic images may decrease over time due to progressive degradation of essential systems characteristics. The TMM (Tissue Mimicking Material) phantom is intended to be used to measure and to enable documentation of changes in void-detectability ratio in periodic tests over years of use. This technical specification establishes:
- Important characteristics and requirements for a TMM 3D artificial cyst phantom using anechoic voids
- A design example of a 3D artificial cyst phantom, the necessary test equipment and use of relevant computer software algorithms.
This technical specification is currently applicable for linear array transducers. A uniformity test prior to void-detectability ratio (VDR) measurement is recommended.

General Information

Status : ACTIVE
Standard Type: Main
Document No: IEC TS 62558:2011
Document Year: 2011
Pages: 49
Edition: 1.0
  • Section Volume:
  • TC 87 Ultrasonics

Life Cycle

Currently Viewing

ACTIVE
IEC TS 62558:2011
Knowledge Corner

Expand Your Knowledge and Unlock Your Learning Potential - Your One-Stop Source for Information!

© Copyright 2024 BSB Edge Private Limited.

Enquire now +