a) deletion of the class designations, class LS/C, class 1/C and class 2/C;
b) addition of two further class designations, class LS/M and class 1/M, specifically for pistonphones;
c) addition of an amended criterion for assessing conformance to a specification: conformance is now demonstrated when (a) measured deviations from design goals do not exceed the applicable acceptance limits and (b) the uncertainty of measurement does not exceed the corresponding maximum-permitted uncertainty;
d) modification to the short-term level fluctuation test of the sound pressure level stability;
e) change to some environmental test conditions to avoid icing;
f) addition of an alternative test for immunity to radio-frequency fields using transverse electromagnetic (TEM) waveguides.">